BCI 1884SS Instrucciones de operaciones Pagina 35

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 43
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 34
Chapter 10: Troubleshooting
Pulse Oximeter for Sleep Screening Operation Manual 10-1
Chapter 10: Troubleshooting
Problem Possible Cause Corrective Action
No pulse shown on the
bargraph.
Patient cable or sensor
is disconnected from the
oximeter.
Sensor is incorrectly
positioned on the
patient.
Poor patient perfusion.
Defective sensor or
patient cable.
Monitor is in the Data
Collection or Trend Print
Mode.
Check sensor connections to the
patient cable and to the oximeter.
Reposition the sensor.
Reposition the sensor.
Try a new sensor or contact your
authorized repair center for help.
None required.
Pulse rate is erratic,
intermittent, or incorrect.
Sensor incorrectly
positioned.
Patient motion
Reposition the sensor.
Patient must remain still to obtain an
accurate measurement.
SpO
2
value is erratic,
intermittent, or incorrect.
Poor patient perfusion.
Patient motion.
Reposition the sensor
Patient must remain still to obtain an
accurate measurement.
The oximeter doesn't
turn on.
Batteries weak.
Batteries not installed or
batteries incorrectly
installed.
Replace the batteries.
Ensure the batteries are installed
correctly.
The oximeter displays
only one flashing LED
with no SpO
2
or heart
rate
The oximeter has been
on continuously and
collecting data for 10
minutes and has gone
into the Data Collection
Mode.
If not collecting trend data, press and
hold the
x
key for 6 seconds, which
will turn off the unit. To return monitor
to Verification Mode, hold
F key for 6
to 8 seconds at power up. This will
clear the trend data and switch the
unit to the Verification Mode.
No printout on built-in
printer.
Batteries are weak.
Batteries are not
properly installed.
No trends in memory.
Printer interface
malfunction.
Replace the batteries.
Ensure the batteries are installed
properly.
Take trend data.
Contact your authorized repair center
for help.
E00 ROM Error Contact service representative
E01 RAM Error Contact service representative