Chapter 12: Specifications
Pulse Oximeter for Sleep Screening Operation Manual 12-1
Chapter 12: Specifications
Equipment Classification
Type of Protection Against
Electric shock:
Internally Powered
Mode of operation: Continuous
Degree of Protection Against
ingress of Liquids:
IPX1, drip proof
Degree of Mobility: Portable
Degree of Protection Against
Electric Shock:
Type BF
Safety Requirements: EN60601-1: 1990
Displays, Indicators, & Keys
SpO2: LED numeric display, 0.43 inches (10.9 mm) high
Pulse Rate: LED numeric display, 0.43 inches (10.9 mm) high
Pulse Strength: Logarithmically scaled 8-segment LED bargraph
Low Battery Indicator: 1 Segment LED
Keys:
x
On/Off key
F
key
SpO
2
Range: 0 - 99% Functional SpO2 (1% increments)
Accuracy: ±2 at 70 - 99% less than 70% is unspecified
Alarms: None
Averaging: 4 pulse beat average
Display Response: The display is to functional saturation.
Display Update Rate: 1 Hz (SpO2); 60 Hz (pulse strength)
Calibration: Factory calibrated over the range of 70% to 100% SpO2
using human blood samples to functional saturation. Test
methods available upon request. No in-service calibration
is required.
Sensor: Red 660nm, 2mW (typical)
Infrared 905nm, 2-2.4mW (typical)