Warnings, Cautions, & Notes
Pulse Oximeter for Sleep Screening Operation Manual v
WARNING! When connecting this monitor to any instrument, verify proper operation
before clinical use. Refer to the instrument’s user manual for full instructions.
Accessory equipment connected to the monitor’s data interface must be certified
according to the respective IEC standards, i.e., IEC 950 for data processing equipment or
IEC 601-1 for electromedical equipment. All combinations of equipment must be in
compliance with IEC 601-1-1 systems requirements. Anyone connecting additional
equipment to the signal input port or the signal output port configures a medical system,
and, therefore, is responsible that the system complies with the requirements of the
system standard IEC 601-1-1.
WARNING! IEC 950 approved equipment must be placed outside the “patient
environment.” The patient environment is defined as an area 1.5 m (4.92 feet) from the
patient.
PATIENT ENVIRONMENT
R 1.5m (4.9 ft)
Cautions
CAUTION! Do not autoclave, ethylene oxide sterilize, or immerse the sensors in liquid.
Evidence that liquid has been allowed to enter the monitor voids the warranty.
CAUTION! This device is intended for use by persons trained in professional health care
or those who have access to the oversight of a professional health care provider. The
operator must be thoroughly familiar with the information in this manual and/or the
“Home User’s Guide” before using the device.
CAUTION! Connect only the PC cable specifically intended for use with this device (see
Optional Supplies and Accessories).
CAUTION! When printing, or advancing the paper, place the oximeter on its’ side to
prevent paper jams.
Notes
NOTE! Operation of this device may be adversely affected in the presence of strong
electromagnetic sources, such as electrosurgery equipment.
NOTE! Operation of this device may be adversely affected in the presence of computed
tomograph (CT) equipment.
NOTE! Use only SpO
2
sensors supplied with, or specifically intended for use with, this
device.
NOTE! SpO
2
measurements may be adversely affected in the presence of high ambient
light. Shield the sensor area (with a towel, for example) if necessary.